9.2 Failed Calibration
If calibration has failed, the GLP function will reset the date and time for the failed parameter to
zero. The
smart
CHEM
-D
still shows the results for the last successful calibration, as shown in the
following example where the last Conductivity calibration has failed…
Cond Zero=0.01uS 00/00/00 00:00
Cond k=1.01 00/00/00 00:00
Cond UnCalibrated
TDS Zero=0.01ppM 31/12/01 12:40
TDS k=1.01 31/12/01 12:50
TDS Calibrated
F2:Back F4:Next
9.3 Printing GLP Information to the RS232 Port
The GLP information stored in the instrument’s memory can be sent to a printer or PC via the
RS232 port.
1. Switch the meter on.
2. Connect one end of the RS232 cable to the
Power/RS232
socket of the
smart
CHEM
-D
.
3. Connect the other end of the RS232 cable to an RS232 Printer, or to a Serial port on a PC.
4. Send the GLP information to the RS232 port:
→
→
F5:GLP
→
→
F3:Print
The message “
Printing GLP Data
” is displayed while sending the data to the RS232 port.
5. The GLP information is sent to the RS232 port in formatted ASCII text. For example…
smartCHEM-D V1.0 S1234 @ 31/12/2001 15:00
Oxygen Zero= 0.1% @ 31/12/2001 12:00
Oxygen Span= 100.0% @ 31/12/2001 12:10
Conductivity Zero= 0.01uS @ 31/12/2001 12:20
Conductivity k= 1.01 @ 31/12/2001 12:30
TDS Zero= 0.01ppM @ 31/12/2001 12:40
TDS k= 1.01 @ 31/12/2001 12:50
Temp. Probe Offset= 1.0oC @ 31/12/2001 14:40
Temp. ED1 Offset= 1.0oC @ 31/12/2001 14:50
Ends