35
9.4.2 How the GLP Alarm Operates
On the day that calibration is due, the following message will be displayed the first time the
smart
CHEM
-D
is switched on…
GLP Alarm, Calibrate
Oxygen
Conductivity/TDS
Temperature
Only those parameters that were selected (see section 9.4.1) will be displayed.
This message will not be displayed again until the next time calibration is due.
In addition to the message shown above, the
smart
CHEM
-D
also flashes the decimal point of each
parameter that requires calibration, and flashes the message “
Cal Now
” on the display.
9.4.3 Notes on the GLP Calibration Alarm function
1. To remove the “
Cal Now
” message and flashing decimal points, each selected parameter must
be correctly calibrated.
2. Switching the GLP Calibration Alarm function off will not clear “
Cal Now
” message and
flashing decimal points. The relevant parameters must be calibrated.
3. When Weekly, 2 Weekly or 4 Weekly periods have been selected, the GLP Calibration Alarm
will always be activated on the same day of the week as when the function was first activated,
even if the unit is not calibrated on that day. For example, if the GLP Calibration Alarm is
triggered weekly on a Monday and the meter is finally calibrated the following Wednesday, the
next GLP Calibration Alarm will still be on the following Monday.
9.5 Instrument Serial Number
In case the serial number that is fitted to the rear of the
smart
CHEM
-D
is removed or becomes
illegible, it is also available on the
smart
CHEM
-D
display.
1. The serial number is displayed at turn-on, for example…
smartCHEM-D V1.0mr
S1234 © 2002 TPS P/L
Oxygen, Conductivity
/TDS, Temperature
The “
m
m
” after
V1.0
is shown when the Dissolved Oxygen Stirrer output is enabled (section 3.2).
The “
r
r
” after
V1.0
is shown when the Recorder Port option is fitted.
2. The serial number is displayed when recalling the GLP information (section 9.1).
3. The serial number is included on the print-out of GLP information (section 9.3).
4. The GLP information can be downloaded to a PC using the optional Windows
software (part
number 130086).
9.6 Additional GLP Features
Another GLP requirement is to record the date and time of every reading. The
smart
CHEM
-D
does
this for you whenever readings are recorded either with the Manual Datalogging function (section
11.1) or the Automatic Logging function (section 11.2).