3.3.2 Span Calibration in Air (all Oxygen modes)
1. Hang the Dissolved Oxygen sensor (and Temperature sensor, if applicable) in air. The tip of the
Dissolved Oxygen sensor should be pointing downwards.
Allow the reading to stabilise. After a zero calibration, this may take up to 5 minutes.
2. Select Oxygen Calibration. (
→
→
F1:Calibrate
→
→
F1:Oxygen
)
When the reading is above approximately 25% Saturation, 2 ppM or 5% Gaseous, the
smart
CHEM
-D
will display the AIR calibration screen…
101*0%S
25.0
O
C
Oxygen AIR Cal.
Press Enter
3. Press
to calibrate. The Automatic Stability Function will now show a and highlight any
unstable readings…
101*0
%S
25.0
O
C
Oxygen AIR Cal.
Waiting to Stabilise
When the Dissolved Oxygen and Temperature readings have both stabilised, the unit will
calibrate itself.
To calibrate immediately without waiting for complete stability, press
to disable the
Automatic Stability Function.
A “
∗
” in the display will be replaced by a decimal point after a successful air calibration.
4. The
smart
CHEM
-D
is now calibrated and is ready for Dissolved Oxygen measurement. Rinse
the Dissolved Oxygen sensor in distilled water and blot dry before placing into unknown
samples.