13. Recorder Port
This section is applicable if the optional Recorder port is fitted.
The optional Recorder Port can be used to send the data from any one parameter to a chart recorder
or other analogue logging device.
13.1 Recorder Port Configuration
1. Select the Recorder set-up menu
(
→
→
F4:Setup
→
→
F4:Ports
→
→
F5:Recorder
).
The
smart
CHEM
-D
will now display the Recorder Port configuration screen…
RECORDER
>F1:Oxygen
F3:Conductivity
F5:Temperature
The arrow indicates the current selection.
2. Press
to set the Recorder output to Dissolved Oxygen data.
Press
to set the Recorder output to Conductivity or TDS data
Press
to set the Recorder output to Temperature data.
Press
to quit and retain the current setting.
13.2 Recorder Port Specifications
Mode
Range
Output Range
Examples
(Reading = mV Out)
Dissolved
Oxygen
0.00 to 40.0 ppM
0.0 to 500 %Sat’n
0.0 to 100 %Gas
0 to 2000 mV
0.00 ppM = 0 mV
100.0 %Sat’n = 400 mV
Conductivity
0 to 3.000 µS/cm
0 to 30.00 µS/cm
0 to 300.0 µS/cm
0 to 3000 µS/cm
0 to 30.00 mS/cm
0 to 300.0 mS/cm
0 to 2000 mV for full
scale of selected
range.
0.000 µS/cm = 0 mV
2760 mS/cm = 1840 mV
TDS
0 to 2.000 ppM
0 to 20.00 ppM
0 to 200.0 ppM
0 to 2000 ppM
0 to 20.00 ppK
0 to 200.0 ppK
0 to 2000 mV for full
scale of selected
range.
0.00 ppM = 0 mV
1000 ppM = 1000 mV
36.0 ppK = 360 mV
Note
: The ranges shown above exceed the range specifications detailed in
section 1.9 to allow for variations due to sensor performance.
Output Impedance
: Approximately 1000 Ohms
Resolution
: Approximately 2 mV