24
5.4 TDS Calibration
Before attempting a TDS calibration, ensure that the
smart
CHEM
-D
has been set up correctly
according to sections 5.1 to 5.3.
Automatic Temperature Compensation is done via a temperature sensor inside the TDS sensor. It is
therefore not necessary to calibrate the Temperature readout of the
smart
CHEM
-D
before taking
TDS measurements.
1. Plug the TDS sensor into the
Cond / TDS
socket.
2. Rinse the TDS sensor in distilled water. Shake off as much water as possible. Blot the outside
of the sensor dry. DO NOT BLOT THE ELECTRODE PLATES.
Zero Calibration
3. Let the sensor dry in air.
4. Select TDS Calibration (
→
→
F1:Calibrate
→
→
F3:TDS
).
5. The
smart
CHEM
-D
will recognise the low TDS signal and attempt a Zero calibration. For
example…
TDS 0*01ppM
ZERO Calibration
Press Enter
6. Press
to calibrate. The Automatic Stability Function will now show a and highlight the
reading while it is unstable…
TDS
0*01
ppM
ZERO Calibration
Waiting to Stabilise
When the TDS reading has stabilised, the unit will calibrate itself.
To calibrate immediately without waiting for complete stability, press
to disable the
Automatic Stability Function.
The “
∗
” will not be removed after a zero calibration.
Continued over the page…