67
18.4 Temperature Troubleshooting
Symptom
Possible Causes
Remedy
Temperature inaccurate and
cannot be calibrated.
1. Faulty connector.
2. Faulty Conductivity/TDS or
Temperature sensor,
whichever is being
calibrated.
Check the connector and replace if
necessary.
Return Conductivity/TDS or Temperature
sensor for repair, or replace sensor.
Displays ‘
OveRnge
’ when
Temperature and/or
Conductivity/TDS sensor
plugged in.
1. Faulty instrument socket.
2. Faulty Conductivity/TDS or
Temperature sensor,
whichever is being
calibrated.
Return the instrument to the TPS factory for
service.
Return Conductivity/TDS or Temperature
sensor for repair, or replace sensor.
Reading does not change in
any sample being measured
Auto Stability Function has
frozen display (notice the ).
Turn Auto Stability Function OFF or re-
start sampling (see section 8).