19
Standard Calibration
7. Place the Conductivity sensor into a sample of Conductivity standard. Ensure that it is
immersed correctly at least to the top of the vent hole in the side of the sensor.
DO NOT
place the sensor directly into the bottle of standard. Discard the used sample of
standard after use.
Select Conductivity Calibration (
→
→
F1:Calibrate
→
→
F1:Conductivity
). The
calibration screen will be displayed with the Conductivity standard to be used. For example…
Cond. 2*86mS/cm
2760.uS/cm
Calibration
Press Enter
8. Press
to calibrate. The Automatic Stability Function will now show a and highlight the
reading while it is unstable…
Cond.
2*86
uS/cm
2760.uS/cm
Calibration
Waiting to Stabilise
When the Conductivity reading has stabilised, the unit will calibrate itself.
To calibrate immediately without waiting for complete stability, press
to disable the
Automatic Stability Function.
The “
∗
” will now be replaced by a decimal point if calibration was successful.
9. The
smart
CHEM
-C
is now calibrated for Conductivity and is ready for use in this mode.