31
Standard Calibration
7. Place the Resistivity sensor into a sample of Resistivity standard. Ensure that it is immersed
correctly at least to the top of the vent hole in the side of the sensor.
DO NOT
place the sensor directly into the bottle of standard. Discard the used sample of
standard after use.
Select Resistivity Calibration (
→
→
F1:Calibrate
→
→
F1:Resistivity
). The
calibration screen will be displayed with the Resistivity standard to be used. For example…
Resist. 352*
Ω
Ω
.cm
362.3
Ω
Ω
.cm
Calibration
Press Enter
8. Press
to calibrate. The Automatic Stability Function will now show a and highlight the
reading while it is unstable…
Resist.
352*
Ω
Ω
.cm
362.3
Ω
Ω
.cm
Calibration
Waiting to Stabilise
When the Resistivity reading has stabilised, the unit will calibrate itself.
To calibrate immediately without waiting for complete stability, press
to disable the
Automatic Stability Function.
The “
∗
” will now be replaced by a decimal point if calibration was successful.
9. The
smart
CHEM
-C
is now calibrated for Resistivity and is ready for use in this mode.