Configuring the EPW 400
OM 100.EPW 400.202002.en
73
6.2.6
Offset limit force sensor
The offset limit is used set the maximum tolerated offset to 10%, 20% or 100% of the
nominal force.
Offset adjustment for sensor with normal signal output:
−
TOX
®
standard sensor: 10% (20% possible for compensation)
By tapping on the input field behind 'Offset limit', the value of the offset limit can
be changed between '10', '20' and '100'. The value displayed last will be adopted.
6.2.7
Forced offset force sensor
If the 'Forced offset' function is activated, the EPW 400 carries out an offset adjust-
ment for the respective channel automatically every time it is switched on. If this func-
tion is not activated, the EPW 400 is ready to measure as soon as it is switched on.
Tap on the input field next to 'Forced offset' to switch the forced offset on or off.
The last setting displayed will be adopted.
6.2.8
Source force sensor
By tapping on the input field behind 'Source', you can switch between normal signal
and DMS. The last source displayed will be adopted.
If DMS is activated, an additional parameter 'Character number' is displayed.
6.2.9
Nominal characteristic value of force sensor
The characteristic for the DMS force transducer is entered using the 'Characteristic
number' parameter.
Value range: 0.1 – 8 mV/V
Tap on the input field next to 'Characteristic number'. The input window containing
the numerical keypad appears.
Enter the characteristic number using the numerical keypad (enter a point before
decimal places; if no decimal places are entered, then the display shows two dec-
imal places '.00').
Confirm the entry with the 'ENTER' button.
Bridge supply voltage
The bridge supply voltage is 5 V.
6.2.10 Filter
With the parameter 'Filter' the cutoff frequency of the measuring channel can be set.
Value range: 5 Hz - 1000 Hz