PLC interface pulse diagrams
OM 100.EPW 400.202002.en
113
7.1.1
Changing the program number
Program bit 0-5
Program strobe
Program ACK
1
Delay time Strobe / Ack: min 10 ms, max. 20 ms
2
New program number applied and data read, max. 100 ms
X
History
I5
I4
I3
I2
I1
I0
Program
0
0
0
0
0
0
0
0
0
0
0
0
1
1
0
0
0
0
1
0
2
0
0
0
0
1
1
3
0
0
0
1
0
0
4
1
1
1
1
0
0
60
1
1
1
1
0
1
61
1
1
1
1
1
0
62
1
1
1
1
1
1
63
7.1.2
Zero point adjustment
Program bit 0
1
Offset external
Program strobe
1
Program ACK
2
Ready for meas-
urement
1
Delay time Strobe / Ack: min 10 ms, max. 20 ms
2
Duration of zero point adjustment: max. 4 sec
Program bit0 = Zero point adjustment either for channel X or channel Y
Program bit0 = 0: X-channel
Program bit0 = 1: Y-channel
Program bit1: If program bit 1 is set, zero point adjustment will take place simulta-
neously for channel X and channel Y
Channel X = 0 / Channel Y = 1