9
Function
(cont.)
Automatic gain
control
(AGC Function)
Receiver gain is automatically adjusted to the optimum level in
response to changes in receiver sensitivity during measurement.
Totalized value
preset
Totalized values can be freely preset.
Preset Range: 0 to 99999999
Error historic
counter
Count "No Echo receiving warning" & "Disturbance detection"
when it occurred.
Power
supply
AC100 to 230V +/-10% (50/60 Hz ±2Hz)
Option: DC24V±20% (This option must be pre-selected.)
Momentary outage
AC input: 20ms
DC input: 0ms
Power
consumption
AC100V: 19VA / AC200V: 23VA
DC24V: 9W (Option)
Fuse
AC: IEC 60127-2 SS5
Cartridge fuse-links,
φ
5.2x20 mm, Rating 2A/250V, Time-lag
and High Breaking Capacity (1500A)
DC: IEC 60127-2 SS5
Cartridge fuse-links,
φ
5.2x20 mm, Rating 4A/250V, Time-lag
and High Breaking Capacity (1500A)
Rush
Current
Less than 15A at AC100V / Less than 25A at AC200V
Less than 15A at DC24V (Option)
Operating
temperature
range
14 to 122°F (-10 to +50°C)
Storage
temperature
range
-4 to 140°F (-20 to +60°C)
Operating
humidity range
Less than 90% RH, non-condensation
Main unit
construction
Protection Degree IP65 (IEC 60529)
Wiring
Connection
port
I/O and power ports: PG13.5 x 3, applicable cable diameter 7 to 12.5 mm
Sensor ports: PG9 x 2, applicable cable diameter 4.5 to 8 mm
Other: USB-B female type for USB communication x 1
Case
material
ABS (Color: White gray)
Weight
Approximately 5 lbs. (2.1 kg)
Dimensions
210mm (W) x 210mm (H) x 100mm (D), not including protrusions
European
Compliance
(CE marking)
EMC Directive 2004/108/EC
Harmonized Standard / EN61326-1:2006 + EN 61326-2-3:2006
Separation into group / Group I
Division into classes / Class A
Location intended for use / In industrial locations
Low Voltage Directive 2006/95/EC
Harmonized Standard / EN61010-1 2nd Edition
Over voltage category II
Pollution degree II
Altitude up to 2000m
Содержание LF810
Страница 1: ...0 PRODUCT DATA SHEET Stationary Ultrasonic Flowmeter LF810 TOSHIBA INTERNATIONAL CORPORATION ...
Страница 13: ...12 7 Dim mension ns M Main Unit D T Dimension Transducer ...
Страница 14: ...13 Mou Mounting unting Fixt Fixture for ure for Tra r Transduc ansducer er Z meth hod ...
Страница 16: ...15 9 System Wiring Connection ...
Страница 18: ...11 Ap Flow ppendix w Volume an x nd Average F Flow Velocity Average e Flow Velocity m s 17 ...