5
4-2. Main Unit
Analog
output
Std / Option
Standard
Number of
channels
1
Output contents
Instantaneous flow rate
Output pattern
Single system
Output format
4 – 20mA, Burnout 20.8mA (when “no echo received” or during
“failure warning” output available)
Max. allowable load resistance 600ohm
Insulated outputs, 10-bit equivalent accuracy
Update cycle
125ms (8Hz)
Terminal panel
Screw less Terminal (0.08~2.5mm
2
cable applicable)
Contact
point output
Std / Option
Standard
Number of
channels
1
Output
Selectable from:
1. Forward flow totalized pulse
2. Backward flow totalized pulse
3. No echo received warning (ROFF)
4. Equipment failure warning
5. No echo received warning or Equipment failure
6. Hi-Limit
alarm
7. Lo-Limit
alarm
8. Forward flow identification
9. Always
open
10. Always closed
Pulse Width
Pulse width of contact is selectable from 1000,500,100 or 20ms.
Note:
1) But not for both forward and backward.
Output format
Photo coupler (insulated)
Contact point
capacity
DC48V, 0.4A
Terminal panel
Screw less Terminal (0.08~2.5mm
2
cable applicable)
USB
communication
Std / Option
Standard
Number of
channels
1
USB Cable
length
Up to 3m
Functions
Flowmeter programming, measurement value display, received
signal waveform display, and log data readout using dedicated
software*1
*1 Compatible with Windows 7, Vista and XP
Connector
USB-B terminal, hot plug possible
Содержание LF810
Страница 1: ...0 PRODUCT DATA SHEET Stationary Ultrasonic Flowmeter LF810 TOSHIBA INTERNATIONAL CORPORATION ...
Страница 13: ...12 7 Dim mension ns M Main Unit D T Dimension Transducer ...
Страница 14: ...13 Mou Mounting unting Fixt Fixture for ure for Tra r Transduc ansducer er Z meth hod ...
Страница 16: ...15 9 System Wiring Connection ...
Страница 18: ...11 Ap Flow ppendix w Volume an x nd Average F Flow Velocity Average e Flow Velocity m s 17 ...