3
1MHz transducer components are as follows.
Components
Quantity
Material
Weight
(approx.)
1. Transducer
1pair (2pcs)
Case : PBT (Plastic)
180g / 2pcs
2. Holder
2pcs
PBT (Plastic)
160g / pc
3. Fastening Fixture
10pcs
SUS304
20g / pc
4. Stainless Band
15 m
SUS430
24g / m
5. Clamp
2sets
SUS304
55g / set
6. Thumb Screw (for Z-method)
4pcs
SUS303
16g / pc
7. Installation Outfit
1set
(30m or less)
Flowmeter cabinet (Pit)
Pipe
Valve
Note: Contact the manufacturer if a pump, valve, pipe with gradual width
increment, or merger pipe is installed at either the upstream or
downstream side.
Power source
Output signal
Conduit pipe (
size
22)
Coaxial cable
Cable trough
(inner diameter = D)
Pump outlet
Pump
Transducer
Cable connection
Flow
(90
bend)
5D or more
Enough length for fully developed
rotationally symmetrical flow profile required.
(Note)
(Note)
Pump
Main unit