The Torrey Pines Scientific HP60 & HS60 Programmable Hot Plates & Hot Plate/Stirrers are 
capable of temperatures in excess of 450

Ε

C at the plate surface.  Touching the heated surface 

will cause severe burns.  

USE EXTREME CAUTION AT ALL TIMES.  

Never leave your hot 

plate accessible to others while it is hot.  Although the unit is equipped with a 

A

HOT 

WARNING

@

 

indicator on the front panel, do not rely on this alone.  It is good practice to never 

touch the heater surface. 

NOTE: DO NOT INSTALL THIS UNIT CLOSER THAN SIX (6) INCHES (15.24 CM) 
TO A WALL OF COMBUSTIBLE MATERIAL.  ALSO, THIS UNIT IS NOT FOR USE 
WITH FLAMMABLE SUBSTANCES. 

 
TEMPERATURE PROBE 
 
When attempting to control PROBE TEMPERATURE, plug in the temperature probe and place 
it in the sample AT ALL TIMES.  If not placed into the sample and plugged into the rear of the 
hot plate, the unit will not be able to sense the temperature of the sample as heat is being applied. 
 This will result in driving the heater to its maximum and could result in ruining the sample. 
 
Temperature probes vary in size and material.  The most commonly used probe is the stainless 
steel 6-inch immersion probe (part number HS30-600).  For other probes available in other 
lengths and materials, call the factory or consult your price list.  All probes come with 3-foot 
(91.44 cm) interconnect cable and are 3/16th (47.6 mm) in diameter.  Stainless steel probes work 
best with all but the most aggressive chemicals.  Solid Teflon will work with the most aggressive 
chemicals, but only to about 260

Ε

C.  Glass probes are available for use with aggressive 

chemicals at higher temperatures. 
 
ELECTRICAL 
 
These hot plate/stirrers are made in models that operate at 100, 115, and 230 volts AC 

10%.  

Be certain that your voltage matches the unit you receive.  Check the nameplate on the bottom 
for the voltage setting on your unit.  Be certain to use a three-wire, properly grounded AC input. 
Take the normal care and precaution one would use with any electrical appliance.  Be careful to 
keep the AC line cord away from the hot plate.  

Be certain to use a line cord of the same type 

and rating as the one supplied with this unit.   Note: all fuses are fast blow.  
Bien vérifier que le cordon utilisé est du même type que celui livré avec l

=

unité.  Note: tous les 

fusibles sont à action rapide.

                           

 
 
 
 
 

 

 

 

 

 

 4

VI. GENERAL DESCRIPTION 

 

The Torrey Pines Scientific HP60/HS60 Series Programmable Hot Plates & Hot Plate/Stirrers 
are general purpose digital/programmable units.  They are available with a solid ceramic top or 

Содержание EchoTherm HP60 SERIES

Страница 1: ...SERIES DOCUMENT NUMBER HP60 HS60 1 Revised 17 January 2011 TORREY PINES SCIENTIFIC INC 2713 Loker Ave West Carlsbad CA 92010 TELEPHONE 760 930 9400 Toll Free 866 573 9104 FAX 760 930 9480 E MAIL info...

Страница 2: ...FRONT AND REAR PANEL CONTROLS 6 7 VIII SET UP INSTRUCTIONS 8 IX DISPLAY AND KEYBOARD DESCRIPTION 9 10 11 X SIMPLE OPERATIONS 11 12 13 XI PROGRAM OPERATIONS 13 14 15 XII TEMPERATURE CALIBRATION 15 16 X...

Страница 3: ...returning items for any reason When applying for authorization please include data regarding the reason the items are to be returned For your protection items must be carefully packed to prevent dama...

Страница 4: ...umber HS30 600 For other probes available in other lengths and materials call the factory or consult your price list All probes come with 3 foot 91 44 cm interconnect cable and are 3 16th 47 6 mm in d...

Страница 5: ...he unit has an AUTO OFF function When the AUTO OFF function is activated the heater and stirrer will turn off at the end of any count down timer setting ALARMS In addition to the audible alarm associa...

Страница 6: ...ral icons which function to display all parameters when the unit is running and during setting When the unit is turned on the display will come on and show the actual plate surface temperature If a pr...

Страница 7: ...The fuse types used appear on the label on the rear of the unit CAUTION If the fuses blow repeatedly contact your dealer or the manufacturer ATTENTION Si les fusibles sautent de mani re r p titive co...

Страница 8: ...container 5 Turn on the unit by the key on the front panel The unit will beep once and the display will light up At this point the user can set or view any of the parameters of the unit 6 Set either t...

Страница 9: ...sition being monitored All stirrer positions can be checked this way Ramp When the RAMP key is depressed the C symbol to the right of the top line of the display with convert to C hr The number value...

Страница 10: ...alue will move around showing each stirrer position Stop on the one to be set and use the UP DOWN RIGHT LEFT arrows to set the stirrer speed in RPM Timer Button The TIMER is for setting the value of t...

Страница 11: ...change to F depress and hold the HEATER OFF button until the display changes To change back to C repeat the process Setting Solution Temperature When setting a solution temperature plug the probe int...

Страница 12: ...that the unit will heat fastest when not using a ramp rate setting even 450 C Hour The ramp rate should only be used to slow the heating rate It is valuable to note that the ramp can be set to go up t...

Страница 13: ...ll be sitting with the same information on the display as when the unit was first turned on To stop a program while it is running touch CANCEL Some rules for programming To erase all programs in memor...

Страница 14: ...display will show 2 for program step two Now enter the values for ramp temperature stirrer and timer being certain to hit ENTER after each Step 2 is now complete 6 Repeat the steps above until the pr...

Страница 15: ...25 C dummy probe in the probe jack at the rear of the unit Depress and hold the HEATER button for 3 seconds until the display reads CAL Then press ENTER Then use the UP DOWN RIGHT LEFT arrows to make...

Страница 16: ...onds until the display reads CAL Then press the ENTER button Use the UP DOWN RIGHT LEFT arrows as needed to make the plate temperature display read what is measured on your meter Then press the ENTER...

Страница 17: ...qu elle est chaude pour viter toute br lure accidentelle Maintenance There is no ongoing maintenance program needed with this unit other than the normal care and cleaning as instructed above and a sim...

Страница 18: ...O Mains Off XV RS232 SERIAL INTERFACE SPECIFICATION See the following pages 18...

Отзывы: