Torrey EchoTherm HP60 SERIES Скачать руководство пользователя страница 10

 
Up/Down, Right/Left Arrows: The UP and DOWN arrows to the right of the display are used to 
set numeric values for the probe temperature, plate temperature, stirrer, ramp, timer, and 
progressing within the steps of a program when writing or editing the program.  Pushing the 
RIGHT and LEFT arrows are for selecting the digit to be set. 
  
Heater button: Touching the HEATER button is for selecting the value to be set into the heater 
plate or probe. Touch HEATER then the RIGHT/LEFT/UP/DOWN arrows to set the target 
temperature  
 
Ramp Button:  Touch the RAMP button is for selecting values to be entered into the ramp 
function. Touch RAMP then the UP/DOWN/RIGHT/LEFT arrows to set the ramp value. To turn 
off the ramp value set it to zero. 
 
Stirrer Button (on units that have stirrers only): Touch the STIRRER button to set the value of 
the stirrer.  Touch STIRRER button then the UP/DOWN/RIGHT/LEFT arrows to set the 
STIRRER value. 
 
Stirrer Select Button (on multi-position stirring units only):  The STIRRER SELECT button is 
for use with the 5-position stirrers only.  It is used with the box to the right of the stirrer numeric 
values to select the stirrer to be controlled.  The stirrers can all be set to one value or each to a 
separate value as needed.  Press STIRRER SELECT over and over and the dot in the box to the 
right of the stirrer value will move around showing each stirrer position.  Stop on the one to be 
set and use the UP/DOWN/RIGHT/LEFT arrows to set the stirrer speed in RPM. 
 
Timer Button: The TIMER is for setting the value of the timer.  Use the 
UP/DOWN/RIGHT/LEFT arrows to select the value to be set into the timer. 
  
Auto-Off Button: Touching and releasing this button activates the AUTO-OFF function.  It is 
used in conjunction with the count-down timer.  When activated, this will shut off the heater and 
stirrer when the timer counts down to zero.  This function can be turned off by touching the 
AUTO-OFF button at any time during a timed event.  When activated, the words 

A

AUTO-OFF

@

 

will be illuminated on the display. 
 
Heater Off Button: Touching and releasing the HEATER OFF button will turn off the heater.  
The HEAT OFF button also is used to change the unit temperature display from 

º

C to 

º

F and 

vise-versa.  Holding the key down for 3 seconds will cause the display to change. 
 
 

 

 

 

 

 

10 

Stirrer Off Button (on units with stirrers only): Touching and releasing the STIRRER OFF 
button will turn off the stirrer.  In the 5-position stirrer this will turn off the stirrer position 
selected with the dot in the box to the right of the stirrer speed display.  Depressing STIRRER 
SELECT will allow the user to scroll around the box to set the dot to the selected stirrer position. 
To turn off all 5 stirrers at the same time, press and hold the STIRRER OFF button until it beeps. 
 

 

 

 

 

 

 

Edit Button: The EDIT button is used to enter the program memory to write or edit a program. 

Содержание EchoTherm HP60 SERIES

Страница 1: ...SERIES DOCUMENT NUMBER HP60 HS60 1 Revised 17 January 2011 TORREY PINES SCIENTIFIC INC 2713 Loker Ave West Carlsbad CA 92010 TELEPHONE 760 930 9400 Toll Free 866 573 9104 FAX 760 930 9480 E MAIL info...

Страница 2: ...FRONT AND REAR PANEL CONTROLS 6 7 VIII SET UP INSTRUCTIONS 8 IX DISPLAY AND KEYBOARD DESCRIPTION 9 10 11 X SIMPLE OPERATIONS 11 12 13 XI PROGRAM OPERATIONS 13 14 15 XII TEMPERATURE CALIBRATION 15 16 X...

Страница 3: ...returning items for any reason When applying for authorization please include data regarding the reason the items are to be returned For your protection items must be carefully packed to prevent dama...

Страница 4: ...umber HS30 600 For other probes available in other lengths and materials call the factory or consult your price list All probes come with 3 foot 91 44 cm interconnect cable and are 3 16th 47 6 mm in d...

Страница 5: ...he unit has an AUTO OFF function When the AUTO OFF function is activated the heater and stirrer will turn off at the end of any count down timer setting ALARMS In addition to the audible alarm associa...

Страница 6: ...ral icons which function to display all parameters when the unit is running and during setting When the unit is turned on the display will come on and show the actual plate surface temperature If a pr...

Страница 7: ...The fuse types used appear on the label on the rear of the unit CAUTION If the fuses blow repeatedly contact your dealer or the manufacturer ATTENTION Si les fusibles sautent de mani re r p titive co...

Страница 8: ...container 5 Turn on the unit by the key on the front panel The unit will beep once and the display will light up At this point the user can set or view any of the parameters of the unit 6 Set either t...

Страница 9: ...sition being monitored All stirrer positions can be checked this way Ramp When the RAMP key is depressed the C symbol to the right of the top line of the display with convert to C hr The number value...

Страница 10: ...alue will move around showing each stirrer position Stop on the one to be set and use the UP DOWN RIGHT LEFT arrows to set the stirrer speed in RPM Timer Button The TIMER is for setting the value of t...

Страница 11: ...change to F depress and hold the HEATER OFF button until the display changes To change back to C repeat the process Setting Solution Temperature When setting a solution temperature plug the probe int...

Страница 12: ...that the unit will heat fastest when not using a ramp rate setting even 450 C Hour The ramp rate should only be used to slow the heating rate It is valuable to note that the ramp can be set to go up t...

Страница 13: ...ll be sitting with the same information on the display as when the unit was first turned on To stop a program while it is running touch CANCEL Some rules for programming To erase all programs in memor...

Страница 14: ...display will show 2 for program step two Now enter the values for ramp temperature stirrer and timer being certain to hit ENTER after each Step 2 is now complete 6 Repeat the steps above until the pr...

Страница 15: ...25 C dummy probe in the probe jack at the rear of the unit Depress and hold the HEATER button for 3 seconds until the display reads CAL Then press ENTER Then use the UP DOWN RIGHT LEFT arrows to make...

Страница 16: ...onds until the display reads CAL Then press the ENTER button Use the UP DOWN RIGHT LEFT arrows as needed to make the plate temperature display read what is measured on your meter Then press the ENTER...

Страница 17: ...qu elle est chaude pour viter toute br lure accidentelle Maintenance There is no ongoing maintenance program needed with this unit other than the normal care and cleaning as instructed above and a sim...

Страница 18: ...O Mains Off XV RS232 SERIAL INTERFACE SPECIFICATION See the following pages 18...