The plate surface calibration procedure cannot be done with the temperature probe 
plugged into the rear of the unit

.  The calibration starts at the low temperatures and goes high. 

 It is time consuming.  Before making an adjustment, be sure that the plate temperature has 
stabilized at the point being calibrated.  That often takes 20 to 30 minutes after the target 
temperature is initially reached.  Follow the procedure below. 
 
1. 

Start at room temperature.  Using your temperature meter, read the plate surface 

 

temperature in the center of the heater plate.  Press and hold the HEATER button for 3 

 

seconds until the display reads CAL. Then press the ENTER button. Use the 

 

UP/DOWN/RIGHT/LEFT arrows to input the temperature measured for the plate 

surface.  

Then press the ENTER button. The plate low calibration point is set. 

 

2.

 

Set the PLATE temperature to 350

Ε

C.  The plate will start to heat.  Give the unit ample 

time to reach and stabilize at 350

Ε

C.  When the temperature is stable, measure the plate 

temperature with your meter.  Press and hold the HEATER button for three seconds until 
the display reads CAL.  Then press the ENTER button.  Use the 
UP/DOWN/RIGHT/LEFT arrows as needed to make the plate temperature display read 
what is measured on your meter.  Then press the ENTER button. 

 
3.

 

Let the unit cool back to room temperature.  It will take some time.  You can speed the 
process by using a fan to help cool the plate.  After about an hour measure the plate 
surface temperature again.  Then reset the plate temperature to what the meter reads as 
was done in step 1 above.   

 
The plate surface calibration is now complete. 

 

 

 
 

 

 

 

 

 

16 

 
 

 
 
 

XIII. CLEANING, MAINTENANCE, AND CONSUMABLE PARTS 

 

Cleaning 

 
These units are subject to splashes and spills during normal use.  Be sure to clean all spills 
quickly.  Wipe spills with a soft cloth or paper towel.  If a cleaning solution is necessary, use a 
mild soap or detergent solution and a soft cloth.  Be sure not to use solvents.  They could damage 
the paint, labels or display window on the unit. A mild abrasive can be used to clean the ceramic 
glass heater surface.  Do not use abrasives on the cast aluminum heater top.  It will scratch the 
surface. 
 

Note:  When using these units with corrosive liquids, the vapors coming off these liquids 
may get into the chassis of the unit and cause damage.  This is particularly true on the 5-

Содержание EchoTherm HP60 SERIES

Страница 1: ...SERIES DOCUMENT NUMBER HP60 HS60 1 Revised 17 January 2011 TORREY PINES SCIENTIFIC INC 2713 Loker Ave West Carlsbad CA 92010 TELEPHONE 760 930 9400 Toll Free 866 573 9104 FAX 760 930 9480 E MAIL info...

Страница 2: ...FRONT AND REAR PANEL CONTROLS 6 7 VIII SET UP INSTRUCTIONS 8 IX DISPLAY AND KEYBOARD DESCRIPTION 9 10 11 X SIMPLE OPERATIONS 11 12 13 XI PROGRAM OPERATIONS 13 14 15 XII TEMPERATURE CALIBRATION 15 16 X...

Страница 3: ...returning items for any reason When applying for authorization please include data regarding the reason the items are to be returned For your protection items must be carefully packed to prevent dama...

Страница 4: ...umber HS30 600 For other probes available in other lengths and materials call the factory or consult your price list All probes come with 3 foot 91 44 cm interconnect cable and are 3 16th 47 6 mm in d...

Страница 5: ...he unit has an AUTO OFF function When the AUTO OFF function is activated the heater and stirrer will turn off at the end of any count down timer setting ALARMS In addition to the audible alarm associa...

Страница 6: ...ral icons which function to display all parameters when the unit is running and during setting When the unit is turned on the display will come on and show the actual plate surface temperature If a pr...

Страница 7: ...The fuse types used appear on the label on the rear of the unit CAUTION If the fuses blow repeatedly contact your dealer or the manufacturer ATTENTION Si les fusibles sautent de mani re r p titive co...

Страница 8: ...container 5 Turn on the unit by the key on the front panel The unit will beep once and the display will light up At this point the user can set or view any of the parameters of the unit 6 Set either t...

Страница 9: ...sition being monitored All stirrer positions can be checked this way Ramp When the RAMP key is depressed the C symbol to the right of the top line of the display with convert to C hr The number value...

Страница 10: ...alue will move around showing each stirrer position Stop on the one to be set and use the UP DOWN RIGHT LEFT arrows to set the stirrer speed in RPM Timer Button The TIMER is for setting the value of t...

Страница 11: ...change to F depress and hold the HEATER OFF button until the display changes To change back to C repeat the process Setting Solution Temperature When setting a solution temperature plug the probe int...

Страница 12: ...that the unit will heat fastest when not using a ramp rate setting even 450 C Hour The ramp rate should only be used to slow the heating rate It is valuable to note that the ramp can be set to go up t...

Страница 13: ...ll be sitting with the same information on the display as when the unit was first turned on To stop a program while it is running touch CANCEL Some rules for programming To erase all programs in memor...

Страница 14: ...display will show 2 for program step two Now enter the values for ramp temperature stirrer and timer being certain to hit ENTER after each Step 2 is now complete 6 Repeat the steps above until the pr...

Страница 15: ...25 C dummy probe in the probe jack at the rear of the unit Depress and hold the HEATER button for 3 seconds until the display reads CAL Then press ENTER Then use the UP DOWN RIGHT LEFT arrows to make...

Страница 16: ...onds until the display reads CAL Then press the ENTER button Use the UP DOWN RIGHT LEFT arrows as needed to make the plate temperature display read what is measured on your meter Then press the ENTER...

Страница 17: ...qu elle est chaude pour viter toute br lure accidentelle Maintenance There is no ongoing maintenance program needed with this unit other than the normal care and cleaning as instructed above and a sim...

Страница 18: ...O Mains Off XV RS232 SERIAL INTERFACE SPECIFICATION See the following pages 18...

Отзывы: