Torrey EchoTherm HP60 SERIES Скачать руководство пользователя страница 11

 
Cancel Button: The Cancel button is used to cancel a program step while writing or editing a 
program or to cancel any setting if made in error. 
 
Run Button: The Run button is used to start running a program in memory. 
 
Enter Button: The Enter button is used to enter a value setting into the unit in manual operation 
or in a program. 
 

 

 

 

 

 

X. SIMPLE OPERATION 

 

The HP/HS60 Series units are fully programmable.  They have the ability to write and store in 
memory ten programs of as many as ten steps each for instant recall and later use.  Also, these 
units can be used to set a single temperature, temperature ramp, stirring speed and timer function 
without going into the memory.  The memory is used for complex routines where all the 
parameters of the sample can be changed automatically to other values without user attention. 

 

 TEMPERATURE 
 
The user may enter either a target probe temperature or a target plate temperature.  Only one 
target temperature is allowed at one time.  Setting a target probe temperature will erase any 
target plate temperature previously set and vice versa. Target temperatures may be set anywhere 
in the range from ambient to 450

Ε

C or 842

º

F in ceramic topped units, or from ambient to 400

º

or 752

º

F in aluminum topped units.  The unit comes with the display reading in degrees C.  To 

change to 

º

F depress and hold the HEATER OFF button until the display changes.  To change 

back to 

º

C repeat the process.   

 
Setting Solution Temperature: When setting a solution temperature, plug the probe into the jack 
on the rear of the unit and place it into the solution.  With the probe plugged in, the unit will only 
set and control the solution temperature as controlled by the probe.  With the probe unplugged, 
only the plate surface temperature is sensed and controlled by the unit.  Next, touch the 
HEATER button and use the UP, DOWN. RIGHT, LEFT arrows to enter the value of the 
temperature wanted.  Then press ENTER.  The unit will now turn the heater on and drive the 
solution to the target temperature.  Note that the temperature display will toggle between the 
target and the actual temperatures.  When the target temperature is reached, the display will still 
toggle but the numerical value will remain the same. 
 

 

 

 

 

 

 

11 

CAUTION: Be certain the temperature probe is in the solution when a solution 
temperature target has been entered.  Failure to do so could damage your sample because 
the hot plate will drive to maximum while seeking a temperature it cannot find.

 

 
PRÉCAUTION D

=

 EMPLOI: Bien vérifier que la sonde de température plonge dans le 

liquide et qu

=

elle est connectée dans la fiche située à l

=

arrière de l

=

appareil lorsque vous 

sélectionnez une température.  L

=

échantillon risque d

=

être endommagé si la sonde n

=

est 

pas connectée car la plaque va monter au maximum de sa température en recherchant une 

Содержание EchoTherm HP60 SERIES

Страница 1: ...SERIES DOCUMENT NUMBER HP60 HS60 1 Revised 17 January 2011 TORREY PINES SCIENTIFIC INC 2713 Loker Ave West Carlsbad CA 92010 TELEPHONE 760 930 9400 Toll Free 866 573 9104 FAX 760 930 9480 E MAIL info...

Страница 2: ...FRONT AND REAR PANEL CONTROLS 6 7 VIII SET UP INSTRUCTIONS 8 IX DISPLAY AND KEYBOARD DESCRIPTION 9 10 11 X SIMPLE OPERATIONS 11 12 13 XI PROGRAM OPERATIONS 13 14 15 XII TEMPERATURE CALIBRATION 15 16 X...

Страница 3: ...returning items for any reason When applying for authorization please include data regarding the reason the items are to be returned For your protection items must be carefully packed to prevent dama...

Страница 4: ...umber HS30 600 For other probes available in other lengths and materials call the factory or consult your price list All probes come with 3 foot 91 44 cm interconnect cable and are 3 16th 47 6 mm in d...

Страница 5: ...he unit has an AUTO OFF function When the AUTO OFF function is activated the heater and stirrer will turn off at the end of any count down timer setting ALARMS In addition to the audible alarm associa...

Страница 6: ...ral icons which function to display all parameters when the unit is running and during setting When the unit is turned on the display will come on and show the actual plate surface temperature If a pr...

Страница 7: ...The fuse types used appear on the label on the rear of the unit CAUTION If the fuses blow repeatedly contact your dealer or the manufacturer ATTENTION Si les fusibles sautent de mani re r p titive co...

Страница 8: ...container 5 Turn on the unit by the key on the front panel The unit will beep once and the display will light up At this point the user can set or view any of the parameters of the unit 6 Set either t...

Страница 9: ...sition being monitored All stirrer positions can be checked this way Ramp When the RAMP key is depressed the C symbol to the right of the top line of the display with convert to C hr The number value...

Страница 10: ...alue will move around showing each stirrer position Stop on the one to be set and use the UP DOWN RIGHT LEFT arrows to set the stirrer speed in RPM Timer Button The TIMER is for setting the value of t...

Страница 11: ...change to F depress and hold the HEATER OFF button until the display changes To change back to C repeat the process Setting Solution Temperature When setting a solution temperature plug the probe int...

Страница 12: ...that the unit will heat fastest when not using a ramp rate setting even 450 C Hour The ramp rate should only be used to slow the heating rate It is valuable to note that the ramp can be set to go up t...

Страница 13: ...ll be sitting with the same information on the display as when the unit was first turned on To stop a program while it is running touch CANCEL Some rules for programming To erase all programs in memor...

Страница 14: ...display will show 2 for program step two Now enter the values for ramp temperature stirrer and timer being certain to hit ENTER after each Step 2 is now complete 6 Repeat the steps above until the pr...

Страница 15: ...25 C dummy probe in the probe jack at the rear of the unit Depress and hold the HEATER button for 3 seconds until the display reads CAL Then press ENTER Then use the UP DOWN RIGHT LEFT arrows to make...

Страница 16: ...onds until the display reads CAL Then press the ENTER button Use the UP DOWN RIGHT LEFT arrows as needed to make the plate temperature display read what is measured on your meter Then press the ENTER...

Страница 17: ...qu elle est chaude pour viter toute br lure accidentelle Maintenance There is no ongoing maintenance program needed with this unit other than the normal care and cleaning as instructed above and a sim...

Страница 18: ...O Mains Off XV RS232 SERIAL INTERFACE SPECIFICATION See the following pages 18...

Отзывы: