program 1 is not the choice use the UP ARROW to scroll to the program of choice. When the
program is chosen, press ENTER and the program stored in that site will start to run. As the
program starts to run, as the unit progresses through the steps in the program, the audible
alarm will sound five times for each step as the program progresses. As the program moves
through the steps in the program the current step will be displayed to the left of the timer.
HOW TO WRITE A PROGRAM
Remember, a program step is a temperature, stirring speed (if wanted), temperature ramp
(if wanted), and a timer setting. Do not advance to the next step until the current one is
complete.
1 Press
EDIT
2
Select the program site (1 through 10) using the UP and DOWN ARROWS.
3
Touch ENTER. The display will show PROG 1 for program one step one.
4
Enter the ramp value if wanted (remember how the timer works with and without a ramp
function) and push the ENTER button. Then enter a temperature and push the ENTER
button, and stirring speed and the ENTER button, and the time function for that step and
ENTER button. This step is now finished.
5
Press the UP ARROW to go to step 2. The display will show 2 for program step two.
Now enter the values for ramp, temperature, stirrer and timer being certain to hit ENTER
after each. Step 2 is now complete.
6
Repeat the steps above until the program is written.
Note: If a mistake is made while entering a program step, just hit CANCEL on that
step and make the new entry. To go back to a prior program step, hit the DOWN
ARROW. Likewise, to go forward to another program step, hit the UP ARROW.
7
After the last program step data has been entered, press ENTER twice. The display will
now show CYC 01. If only one cycle of the program is needed, hit ENTER. If more
than one cycle of the program is wanted, use the UP ARROW to go to the number of
cycles wanted. For example, twenty cycles would appear on the display as CYC 20.
Then press ENTER again and the program is written and stored for use.
14
8
To run a program, press RUN, the display will show PROG 1. To run program 1, press
ENTER. To select and run a different program, use the UP ARROW to scroll to the
program to be run, and press ENTER.
XII. TEMPERATURE CALIBRATION
The HP/HS60 series are designed for accuracy. The temperature calibration designed into the
unit is made to hold for very long periods of time. When calibrated in the factory, it is expected
that the unit will meet the most demanding customers
=
requirements. However, our standards