Torrey EchoTherm HP60 SERIES Скачать руководство пользователя страница 14

    program 1 is not the choice use the UP ARROW to scroll to the program of choice.  When the 
    program is chosen, press ENTER and the program stored in that site will start to run.   As the  
    program starts to run, as the unit progresses through the steps in the program, the audible         
    alarm will sound five times for each step as the program progresses.  As the program moves    
    through the steps in the program the current step will be displayed to the left of the timer. 
 
HOW TO WRITE A PROGRAM 
 

Remember, a program step is a temperature, stirring speed (if wanted), temperature ramp 
(if wanted), and a timer setting.  Do not advance to the next step until the current one is 
complete.

 

 
1 Press 

EDIT 

Select the program site (1 through 10) using the UP and DOWN ARROWS. 

Touch ENTER.  The display will show PROG 1 for program one step one. 

Enter the ramp value if wanted (remember how the timer works with and without a ramp 
function) and push the ENTER button.  Then enter a temperature and push the ENTER 
button, and stirring speed and the ENTER button, and the time function for that step and 
ENTER button. This step is now finished. 

Press the UP ARROW to go to step 2.  The display will show 2 for program step two.  
Now enter the values for ramp, temperature, stirrer and timer being certain to hit ENTER 
after each.  Step 2 is now complete. 

Repeat the steps above until the program is written. 

 

Note: If a mistake is made while entering a program step, just hit CANCEL on that 
step and make the new entry.  To go back to a prior program step, hit the DOWN 
ARROW.  Likewise, to go forward to another program step, hit the UP ARROW. 

 

After the last program step data has been entered, press ENTER twice.  The display will 
now show CYC 01.  If only one cycle of the program is needed, hit ENTER.  If more 
than one cycle of the program is wanted, use the UP ARROW to go to the number of 
cycles wanted.  For example, twenty cycles would appear on the display as CYC 20.  
Then press ENTER again and the program is written and stored for use. 

 

 

 

 

 

 

14 

 

To run a program, press RUN, the display will show PROG 1.  To run program 1, press 
ENTER.  To select and run a different program, use the UP ARROW to scroll to the 
program to be run, and press ENTER.   

      

 

 

 

XII. TEMPERATURE CALIBRATION 

 

 
The HP/HS60 series are designed for accuracy.  The temperature calibration designed into the 
unit is made to hold for very long periods of time.  When calibrated in the factory, it is expected 
that the unit will meet the most demanding customers

=

 requirements.  However, our standards 

Содержание EchoTherm HP60 SERIES

Страница 1: ...SERIES DOCUMENT NUMBER HP60 HS60 1 Revised 17 January 2011 TORREY PINES SCIENTIFIC INC 2713 Loker Ave West Carlsbad CA 92010 TELEPHONE 760 930 9400 Toll Free 866 573 9104 FAX 760 930 9480 E MAIL info...

Страница 2: ...FRONT AND REAR PANEL CONTROLS 6 7 VIII SET UP INSTRUCTIONS 8 IX DISPLAY AND KEYBOARD DESCRIPTION 9 10 11 X SIMPLE OPERATIONS 11 12 13 XI PROGRAM OPERATIONS 13 14 15 XII TEMPERATURE CALIBRATION 15 16 X...

Страница 3: ...returning items for any reason When applying for authorization please include data regarding the reason the items are to be returned For your protection items must be carefully packed to prevent dama...

Страница 4: ...umber HS30 600 For other probes available in other lengths and materials call the factory or consult your price list All probes come with 3 foot 91 44 cm interconnect cable and are 3 16th 47 6 mm in d...

Страница 5: ...he unit has an AUTO OFF function When the AUTO OFF function is activated the heater and stirrer will turn off at the end of any count down timer setting ALARMS In addition to the audible alarm associa...

Страница 6: ...ral icons which function to display all parameters when the unit is running and during setting When the unit is turned on the display will come on and show the actual plate surface temperature If a pr...

Страница 7: ...The fuse types used appear on the label on the rear of the unit CAUTION If the fuses blow repeatedly contact your dealer or the manufacturer ATTENTION Si les fusibles sautent de mani re r p titive co...

Страница 8: ...container 5 Turn on the unit by the key on the front panel The unit will beep once and the display will light up At this point the user can set or view any of the parameters of the unit 6 Set either t...

Страница 9: ...sition being monitored All stirrer positions can be checked this way Ramp When the RAMP key is depressed the C symbol to the right of the top line of the display with convert to C hr The number value...

Страница 10: ...alue will move around showing each stirrer position Stop on the one to be set and use the UP DOWN RIGHT LEFT arrows to set the stirrer speed in RPM Timer Button The TIMER is for setting the value of t...

Страница 11: ...change to F depress and hold the HEATER OFF button until the display changes To change back to C repeat the process Setting Solution Temperature When setting a solution temperature plug the probe int...

Страница 12: ...that the unit will heat fastest when not using a ramp rate setting even 450 C Hour The ramp rate should only be used to slow the heating rate It is valuable to note that the ramp can be set to go up t...

Страница 13: ...ll be sitting with the same information on the display as when the unit was first turned on To stop a program while it is running touch CANCEL Some rules for programming To erase all programs in memor...

Страница 14: ...display will show 2 for program step two Now enter the values for ramp temperature stirrer and timer being certain to hit ENTER after each Step 2 is now complete 6 Repeat the steps above until the pr...

Страница 15: ...25 C dummy probe in the probe jack at the rear of the unit Depress and hold the HEATER button for 3 seconds until the display reads CAL Then press ENTER Then use the UP DOWN RIGHT LEFT arrows to make...

Страница 16: ...onds until the display reads CAL Then press the ENTER button Use the UP DOWN RIGHT LEFT arrows as needed to make the plate temperature display read what is measured on your meter Then press the ENTER...

Страница 17: ...qu elle est chaude pour viter toute br lure accidentelle Maintenance There is no ongoing maintenance program needed with this unit other than the normal care and cleaning as instructed above and a sim...

Страница 18: ...O Mains Off XV RS232 SERIAL INTERFACE SPECIFICATION See the following pages 18...

Отзывы: