38
3
When the beam is positioned so the two measurements are the same, press the [ENTER] key.
The [V] LED will flash.
When the flashing stops, the vertical calibration adjustment is made and power is turned off.
Repeat the checking procedure to confirm proper calibration has been made.
This operation can be performed with the remote controller RC-40.
When using the RC-40, press the [ENTER] key on the RC-40 unit.
(The [ENTER] key for the RC-40: The same mark as the [ENTER] key for the
instrument.)
NOTE
If the calibration is greater than the adjustment allows, the error LED will start
flashing. If this occurs, contact your Topcon dealer.