29
Checking and Adjusting
There are three areas of performance the user should check periodically.
Horizontal Calibration
Horizontal Rotation Cone
Vertical Calibration
The Horizontal Calibration and Vertical Calibration can be easily checked and, in most cases,
adjustments can be made by the user. Horizontal Rotation Cone can be checked by the user, but if
an error is found, adjustments must be made by a Topcon service facility.
Attaching the calibration decals
Before calibration, attach the calibration decals to the instrument as shown below. The calibration
decal shows the calibration function of certain control keys on the control panel.
CALIBRATION
START-While pressing ,press
X
CAL
-ON,Y
CAL
-press ON
V
CAL
-ON
LASER BEAM CORRECTION
- or
ENTER-press
START
1
1
X
V
Y
2
3
ENTER
ERROR
4
2
3
4
X / Y
X / Y
X
Y
V
Calibration Decals
Содержание RL-VH4DR/G2
Страница 1: ...INSTRUCTION MANUAL ROTATING LASER RL VH4DR G2 31375 90050 TEP...
Страница 2: ...EU...
Страница 47: ...45...
Страница 55: ...2008 TOPCON CORPORATION ALL RIGHTS RESERVED...