32
(2) Adjusting Calibration
If the distance between either set of marks is more than 6 mm (1/4 of an inch) but less than
25 mm (1 inch), turn the unit off by pressing the [START] key once and use the following
procedure to calibrate the laser. Confirm that unit has shut off before beginning the procedure.
(In steps 2 and 3, use of optional RC-40 remote controller can be helpful. See page 34.)
1
While pressing the [ENTER] key, press the [START] key. This activates the X axis calibration
mode. Confirm that the [X] LED is lit.
2
By pressing the right or left Alignment Control key, move the X2 (Mb) laser beam up or down
until its centered between marks Ma and Mb.
NOTE
• This operation can be performed with the Remote Controller RC-40.
• When using the RC-40, select the X axis with the X/Y axis selection key and move the
laser beam up or down with the alignment key.
START
1
X
V
Y
2
3
ENTER
ERROR
4
X / Y
Laser beam
up or down
Center
position