6-6
6 SPECIAL MODE (Menu Mode)
5
Press the [F1] or [F2] key to select using
GRID FACTOR.
[Example: F2: DON’T USE]
[F2]
6
Press the [F1] key.
[F1]
7
Collimate prism A, and press the [F1](MEAS) key.
Horizontal distance (HD) between the instrument
and prism A will be shown.
Collimate A
[F1]
8
After measuring, the prism position will be
decided.
9
Collimate prism B and press the [F1](MEAS) key.
Horizontal distance (HD) between the instrument
and prism B will be shown.
Collimate B
[F1]
10
After measuring, the horizontal distance (dHD)
and relative elevation (dVD) between prism A and
B will be shown.
[F4]
11
To show slope distance (dSD), press [
] key.
[
]
12
To measure the distance between points A and C,
press the [F3](HD). *1)
[F3]
13
Collimate point C (Prism C) and press the
[F1](MEAS) key.
Horizontal distance (HD) between the instrument
and prism C will be shown.
Collimate
prism C
[F1]
14
After measuring, the horizontal distance (dHD)
and relative elevation (dVD) between prism A and
C will be shown.
[F4]
15
To measure the distance between points A and D,
repeat procedure 12 to14. *1)
*1) To return to previous mode, press the [ESC] key.
MLM
F1:MLM-1(A-B, A-C)
F2:MLM-2(A-B, B-C)
MLM-1(A-B, A-C)
<STEP-1>
HD: m
MEAS R.HT NEZ –––
MLM-1(A-B, A-C)
<STEP-1>
HD*[n] << m
>Measuring...
MLM-1(A-B, A-C)
<STEP-2>
HD: m
MEAS R.HT NEZ –––
MLM-1(A-B, A-C)
<STEP-2>
HD* [n] << m
MEAS R.HT NEZ –––
MLM-1(A-B, A-C)
dHD : 123.456 m
dVD : 12.345 m
––– ––– HD –––
MLM-1(A-B, A-C)
dSD : 234.567 m
HR : 12°34'40"
––– ––– HD –––
MLM-1(A-B, A-C)
<STEP-2>
HD: m
MEAS R.HT NEZ –––
MLM-1(A-B, A-C)
dHD : 234.567 m
dVD : 23.456 m
––– ––– HD –––