6-17
6. Troubleshooting
z
z
z
z
z
TOTAL, CONTRAST volumes cannot be used.
Cause 1
The volume controller is not switched as TOTAL /
CONTRAST.
Remedy
Press the Volume Function Select button-1
to select it to TOTAL / CONTRAST.
z
z
z
z
z
NEAR / FAR volumes cannot be used.
Cause 1
The volume controller is not swithced as NEAR /
FAR.
Remedy
Press the Volume Function select button-2
to select it to NEAR / FAR.
Cause 2
The images are frozen.
Remedy
NEAR / FAR Gain control works only on scan-
ning the images. Take pictures to use NEAR
/ FAR Gain control. Old captured images will
be lost, after starting taking pictures.
z
z
z
z
z
"Probe Error!!" comes up on screen when turning the unit on or
when switching to B-scan mode.
Cause 1
The probe is not properly connected.
Remedy
Make sure that the probe is inserted suffi-
ciently.