5-3
5.3 Routine maintenance
b) Cleaning
For cleaning the eye contacting part of the probe, use
absorbent cotton to avoid damaging the part.
Do not pull the cord with an undue force, which may
otherwise cause power disconnection.
The connector jointed part shall not be wetted by clean-
ing.
z
For cleaning the probe, wipe the head with an absorbent cot-
ton soaked with ethanol for disinfection and then clean the
head with water.
z
Be sure to sufficiently remove water after cleaning is given.
c) Disinfection
No sterilization by autoclave or EOG is allowed, which
may otherwise greatly damage the probe.
z
Sterilization of the B-scan probe, the A-scan probe, and
the Biometry probe shall be performed by first immersing
the probe up to within 15mm from the eye contacting part
ethanol for disinfection or sodium hypochlorite acid sol-
vent (0.5%) for 10 to 20 minutes and then clean the probe
with purified water.
Within 15mm
Within 15mm
Содержание UD-600
Страница 2: ......
Страница 14: ...xii This page is intentionally blank...
Страница 22: ...1 8 This page is intentionally blank...
Страница 28: ...2 6 This page is intentionally blank...
Страница 157: ...3 129 Simple mode 3 9 Printout...
Страница 159: ...3 131 Simple mode 3 9 Printout...
Страница 192: ...3 164 This page is intentionally blank...
Страница 256: ...8 14 This page is intentionally blank...
Страница 260: ...A 4 INDEX...
Страница 262: ...0123 ID 0608...