![Tomey UD-600 Скачать руководство пользователя страница 222](http://html1.mh-extra.com/html/tomey/ud-600/ud-600_operators-manual_1125672222.webp)
6-8
6. Troubleshooting
<Patient data inquiry to the TOMEY Link>
z
z
z
z
z
"Comm. Busy!! Retry!!" is displayed.
Cause 1
Because of the internal communications, communi-
cations are not received temporarily.
Remedy
Give operation again after a short time of
elapse.
z
z
z
z
z
"NO ID!!" is displayed.
Cause 1
Patient ID number has not been inputted.
Remedy
Give communication operation after inputting
the patient ID number.
z
z
z
z
z
"LAN Connection ERROR!!" is displayed.
Cause 1
The connection of the LAN cable is not correct.
Remedy
Check to see if the connection of the LAN
cable to the connector provided at the rear
side of the instrument unit and to the net-
work side (the switching hub) is proper. Es-
pecially as for the connection of the LAN
cable to the network side, ask your system
administrator for correct connection.
Cause 2
The power of the hub is not on.
Remedy
Check if the power of the hub wher LAN cable
is connected is on.
z
z
z
z
z
"Server Timeout!!" is displayed.
Cause 1
No server's response is caused by busy condition
of the TOMEY Link server.
Remedy
Give operation again after a short time of
elapse.
Содержание UD-600
Страница 2: ......
Страница 14: ...xii This page is intentionally blank...
Страница 22: ...1 8 This page is intentionally blank...
Страница 28: ...2 6 This page is intentionally blank...
Страница 157: ...3 129 Simple mode 3 9 Printout...
Страница 159: ...3 131 Simple mode 3 9 Printout...
Страница 192: ...3 164 This page is intentionally blank...
Страница 256: ...8 14 This page is intentionally blank...
Страница 260: ...A 4 INDEX...
Страница 262: ...0123 ID 0608...