ZT-1 User’s Manual – Operation
DOC #18172 Page 27
7) Adjust the gain.
•
Refer to calibration data report for specific “cal. number” in VDC. If no calibration
data report, apply known load.
•
Adjust the GAIN pot until the digital volt meter reads same as “cal. Number” VDC.
If no calibration data report, adjust gain pot to desired output voltage. eg. 7.0 VDC.
•
Repeat steps 5-7 until the correct voltages are achieved.
For example, +7.0V. 10.0 VDC is the maximum, verifiable with a calibrated
multimeter.
Note: Use the VOLT LED in the built-in bar-graph voltmeter as an indicator of
movement to required cal. number value. This characteristic applies to all LED’s in
this built-in bar-graph voltmeter.