ZT-1 User’s Manual – Operation
DOC #18172 Page 28
8) Record the Calibration Gain Voltage.
•
Set the mode switches to AZ OFF, PRB OFF, TRACK, and ZERO as shown,
and verify that the sensor is at zero.
•
Set the ZRO/GAIN switch to GAIN and use a digital voltmeter to measure the
voltage across the OUT and COM terminals. Record this value as the
Calibration Gain Voltage.
Figure 13:
Calibration Gain Voltage Measurement