USER MANUAL SAGA
Page 41 of 47
Guidance and manufacturer’s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should
assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test level
Complian
ce level
Electromagnetic environment – guidance
Voltage dips
IEC 61000-4-
11
0 % U
T 2
; ½ cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0 % U
T 2
; 1 cycle
and
70 % U
T 2
; 25/30
3
cycles
Single phase: at 0°
0 % U
T
0 % U
T
Mains power quality should be that of a typical commercial or
hospital environment.
Voltage
interruptions
IEC 61000-4-
11
0 % U
T 2
; 250/300
3
cycles
0 % U
T
Degradation of performance may occur: Interruption of AC
mains may require measurement restart. For further
information refer to chapter 8.1.
If the user of the product requires continued operation during
power mains interruptions, it is recommended that the product
be powered from an uninterruptible power supply or a battery.
Electrical
transient
conduction
along supply
lines
ISO 7637-2
As specified in ISO 7637-2
N.A.
6
Notes:
1.
This is the R.M.S. level, before modulation is applied.
2.
U
T
is the A.C. mains voltage prior to application of the test level.
3.
E.g. 10/12 means 10 periods at 50 Hz or 12 periods at 60 Hz.
5.
Applies only to devices with magnetically sensitive components or circuitry.
6.
Applies only to devices intended to be powered from 12/24 V road vehicle power systems.