Guidance and manufacturer’s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should
assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test level
Complian
ce level
Electromagnetic environment – guidance
Proximity fields
from RF
wireless
communication
s equipment
IEC 61000-4-3
27 V/m
At 385 MHz, 1.8 W,
18 Hz pulse modulation
28 V/m
At 450 MHz, 2 W,
1 kHz sine FM modulation with
±5 kHz deviation
9 V/m
At 710 MHz, 745 MHz and
780 MHz, 0.2 W,
217 Hz pulse modulation
28 V/m
At 810 MHz, 870 MHz and
930 MHz, 2 W,
18 Hz pulse modulation
28 V/m
At 1720 MHz, 1845 MHz and
1970 MHz, 2 W,
217 Hz pulse modulation
28 V/m
At 2450 MHz, 2 W,
217 Hz pulse modulation
9 V/m
At 5240 MHz, 5500 MHz and
5785 MHz, 0.2 W,
217 Hz pulse modulation
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
28 V/m
9 V/m
Degradation of performance may occur: Electromagnetic
radiation may distort and disturb acquired signals. At some
frequencies the live communication link may be interrupted.
For further information refer to chapter 8.1.
Portable and mobile RF communications equipment should be
used no closer to any part of the product, including cables,
than 0.3 m.
Electrical fast
transients /
burst
IEC 61000-4-4
±2 kV
100 kHz repetition frequency
for power supply lines
±1 kV
for input/output lines
±2 kV
±1 kV
Degradation of performance may occur: Electrical transient
pulses on AC mains and USB connection may cause signal
loss and require a measurement restart in wireless operation.
In docked operation transient signal interruptions during pulses
may occur. For further information refer to chapter 8.1.
Mains power quality should be that of a typical commercial or
hospital environment.
Surges
IEC 61000-4-5
±1 kV
Line-to-line
±2 kV
Line-to-ground
±1 kV
±2 kV
Degradation of performance may occur: During surges
transient signal interruptions may occur. For further information
refer to chapter 8.1.
Mains power quality should be that of a typical commercial or
hospital environment.
Conducted
disturbances
induced by RF
fields
IEC 61000-4-6
3 V
1
(0.15 to 80) MHz
80 % AM at 1 kHz
6 V
1
in ISM and amateur radio
bands within (0.15 to 80) MHz
80 % AM at 1 kHz
3 V
6 V
Degradation of performance may occur: High level conducted
disturbances on AC mains may cause data communication
loss, on other connections including subject connections they
may cause transient signal interruptions. For further
information refer to chapter 8.1.
Rated power
frequency
magnetic fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
N.A.
5
The product is not intended for environments with restricted
access due to strong magnetic fields.