Electromagnetic emission
Guidance and manufacturer’s declaration – electromagnetic emissions
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product
should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The product uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
(Docking Station)
Class B
(Data Recorder)
NOTE The EMISSIONS characteristics of the
Docking Station make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used
in a residential environment (for which CISPR 11
class B is normally required) it might not offer
adequate protection to radio-frequency
communication services. The user might need to
take mitigation measures, such as relocating or re-
orienting it.
The Data Recorder is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / flicker
emissions
IEC 61000-3-3
Complies
Electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should
assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test level
Complian
ce level
Electromagnetic environment – guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV
contact
±15 kV
air
±8 kV
±15 kV
Degradation of performance may occur: Electrostatic
discharges may result in air sparks at connectors coincident
with signal loss. Sometimes a measurement restart is required.
For further information refer to chapter 8.1.
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %.
Radiated RF
EM fields
IEC 61000-4-3
10 V/m
(80 to 2700) MHz
80 % AM at 1 kHz
10 V/m
Degradation of performance may occur: Electromagnetic
radiation may distort and disturb acquired signals. At some
frequencies the live communication link may be interrupted.
For further information refer to chapter 8.1.