Push-Button…
… behind the slot
for battery cover
Reset
Should strong electromagnetic fields cause a malfunction of your VD82EX, you
can perform a reset of your gauge using the push-button in the battery
compartment:
1. Push the battery cover downwards and open the battery compartment.
2. Press push-button by means of a small screwdriver or other suitable tool.
3. Close battery cover.
Error messages and malfunction
Problem
Possible Cause
Correction
high measurement error
contamination, ageing,
extreme temperature,
maladjustment
readjustment
display shows "or"
pressure over range
(pressure > 1050mbar)
error message "Err"
adjustment done at wrong
pressure
measurement error out of
adjustment range
displayed pressure must be
>800mbar f. atmosphere
adjustment, <100mbar f.
zero adjustment
send unit for repair
error message "Err1"
defective sensor
send unit for repair