MODE
5s
5 Configuration
To switch the VD82EX into Configuration-Mode:
with the instrument switched-off hold the
Mode-Key pressed for approx. 5 seconds,
until the display shows ‘’Stor’’.
5.1 Memory Function
To activate the memory function of your VD82EX for saving minimum and
maximum pressure values, switch the unit into Configuration-Mode. For this the
instrument must be switched-off. Hold the Mode-Key pressed then, until the
display shows "Stor":
After 5 more seconds the current status of the memory function is displayed:
With the Mode-Key the memory function can
now be enabled or disabled.
Without further keystroke, the unit switches to Auto-Off-Mode after approx. 5
seconds. The last settings are saved.
5.2 Adjustment
The instrument is factory adjusted. Through use under different climatic
conditions, through extreme temperature changes, ageing or contamination
readjustment can become necessary.
To adjust your VD82EX, switch the unit into Configuration-Mode. For this the
instrument must be switched-off. Hold the Mode-Key pressed then, until the
display shows "Stor".
Then press Mode-Key several times, until the display shows "CAL":
MODE
5s
MODE
MODE