30
Adjustment on Atmosphere Pressure
After 5 more seconds the display shows:
Press Mode-Key, the actual atmosphere pressure is displayed:
Using the
Mode-Key you can now adjust the
reference pressure: very keystroke changes
the displayed value for another 1mbar alter-
nating up- and downwards.
After approx. 5 seconds without further keystroke, the last setting is taken over
and adjustment procedure finished. The unit switches to Auto-Off-Mode.
Adjustment on atmosphere pressure is possible only if the
displayed actual pressure is above 800 mbar. Otherwise
adjustment is denied and the error message "Err" displayed.
Adjustment on Zero Pressure
For adjustment on zero pressure the actual pressure inside the
sensor has to be less than 0,1 mbar !
The displayed pressure must be less than 100 mbar,
otherwise adjustment is denied and the error message "Err"
displayed..
Switch the unit to Configuration-Mode like described above and press Mode-Key
several times, until "CAL" is displayed.
After 5 more seconds the display shows:
After 5 more seconds the display shows:
5s
5s
MODE
mbar
MODE
mbar