4.7 LDC340 specific commands
LDC340 / Page 46
4.7.8 Laser diode current (ILD)
Programming:
Syntax:
"
:ILD:SET <NR3>
"
Valid Range: Depends on the instrument type.
Default Value: 0 A
Assumption: The operation mode is switched to constant current.
Description:
Sets the laser diode current. Unit: [A].
Reading the set current:
Syntax:
"
:ILD:SET?
"
Response:
[:ILD:SET <NR3>]
Description:
Queries the programmed laser diode current. Unit: [A].
Reading the minimum current:
Syntax:
"
:ILD:MIN?
"
Response:
[:ILD:MIN <NR3>]
Description:
Queries the minimum allowed laser diode current. Unit: [A].
Reading the maximum current:
Syntax:
"
:ILD:MAX?
"
Response:
[:ILD:MAX <NR3>]
Description:
Queries the maximum allowed laser diode current. Unit: [A].
Reading the actual current:
Syntax:
"
:ILD:ACT?
"
Response:
[:ILD:ACT <NR3>]
Description:
Queries the actual laser diode current. Unit: [A].