6.3 Technical data
LDC340 / Page 77
Measurement resolution photo current 1 (remote control)
0.01
µ
A / 0.1
µ
A
Accuracy of photo current 1
±
2 µA / ± 20 µA
Display optical power 1
0 ... 2 W
Resolution optical power 1
100
µ
W
Bias voltage photo diode 1
0 ... 10 V
Control range of photo current 2
5
µ
A ... 2 mA
Setting resolution photo current 2 (manual control)
0.1
µ
A
Setting resolution photo current 2 (remote control)
0.03
µ
A
Measurement resolution photo current 2 (manual control)
0.1
µ
A
Measurement resolution photo current 2 (remote control)
0.01
µ
A
Accuracy of photo current 2
±
2 µA
Bias voltage photodiode 2
n.a.
LD Current limit
Setting range
0 ...
≥
4 A
Measurement resolution (manual / remote control)
1 mA / 0.1 mA
Accuracy
±
10 mA
PD Current limit
Setting range
0 ...
≥
2 mA / 10 mA
Measurement resolution (manual control)
0.1 µA / 1 µA
Measurement resolution (remote control)
10 nA / 100 nA
Accuracy
±
20 µA / ± 200 µA
Laser voltage measurement
Measurement principle
4-wire
Measurement range
0 ... 10 V
Measurement resolution (manual / remote control)
1 mV / 100µV
Accuracy
±
10 mV
Analog modulation input (MOD IN)
Input impedance
10 k
Ω
Max. input voltage
±10 V
Small signal 3 dB bandwidth (CC)
DC ... 50 kHz
Laser diode modulation coefficient (CC)
400 mA/V ± 5%
Laser diode modulation coefficient (CP)
0.2 mA/V ± 5% / 2 mA/V ± 5%