3.4 Main setting knob
LDC340 / Page 20
WIN
lights up if the temperature window function of a TED 350 connected
to the LDC340 via BNC cable protects the laser from being switched
on.
OTP
lights up if the internal temperature of the LDC340 is too high. Please
wait until the LED extinguishes. Then the LDC340 can be operated
again.
INTERLOCK
lights up if the interlock loop is open.
3.4 Main setting knob
By means of this 10-turn setting knob the set values or actual values for laser
current, monitor current or optical power are set according to operating mode and
status.
If the laser current is switched off the LED “PRESET“ will light up. The set value of
the selected parameter is displayed.
If the laser current is switched on the LED “PRESET“ will remain off. The actual value
of the selected parameter is displayed.
3.5 Mode Select
Four modes of operation can be selected by consecutively pressing the "mode
select" key:
•
Constant current mode in the high current range (version dependent)
•
Constant current mode in the low current range (version dependent)
•
Constant power mode using an internal monitor diode
•
Constant power mode using an external monitor diode
A corresponding LED indicates the selected mode.