4.8 Error messages of the LDC340-IEEE
LDC340 / Page 55
4.8.2 LDC340 operation error messages
[
1001,"Interlock is open"
]
Category:
Execution
Error
Possible reason:
Try to switch on the output while the interlock line is open.
(Refer to chapter 3.2.2, "Using the interlock input" on page 16)
[
1002,"Open circuit"
]
Category:
Execution
Error
Possible reason:
Try to switch on the output while the cable to the laser diode has
opened or the compliance voltage is not high enough.
[
1003,"Over temperature"
]
Category:
Execution
Error
Possible reason:
Try to switch on the output while the internal temperature is too
high. Wait until the LDC340 has cooled down. Maintain proper air
flow.
[
1005,"No LD polarity change during laser on"
]
Category:
Execution
Error
Possible reason:
The polarity of the laser diode can not be changed while the laser
diode output is switched on.
[
1006,"No PD polarity change during laser on"
]
Category:
Execution
Error
Possible reason:
The polarity of the monitor diode can not be changed while the
laser diode output is switched on.
[
1007,"No setting of ILD during constant power mode"
]
Category:
Execution
Error
Possible reason:
The set value of the laser diode current can not be changed
during constant power mode.