![THORLABS LDC340 Скачать руководство пользователя страница 73](http://html.mh-extra.com/html/thorlabs/ldc340/ldc340_operation-manual_1110763073.webp)
5.1 Maintenance
LDC340 / Page 67
5 Maintenance and repair
5.1 Maintenance
Protect the LDC340 from adverse weather conditions. The LDC340 is not water
resistant.
Attention
To avoid damage to the LDC340, do not expose it to spray, liquids or
solvents and do not use any abrasive or chemical cleaning agents.!
The LDC340 does not need a regular maintenance.
If necessary the unit and the LCD display can be cleaned with a cloth dampened with
water.
You can use a mild 75% Isopropyl Alcohol solution for more efficient cleaning.
To guarantee the specifications given in chapter 6.3 over a long period it is
recommended to have the unit calibrated by
Thorlabs GmbH
every two years.
There are no user serviceable modules inside. Any service must only be done by
qualified service personnel.
The LDC340 does not contains any components to be repaired by the user. If any
disturbances in function occur please first contact the
Thorlabs GmbH
technical
hotline (refer to section 6.7, "Addresses" on page 83) before sending the device to
Thorlabs GmbH
for repair.