SD-OCT Base Unit
Chapter 7: Troubleshooting
Page 56
MTN020880-D02
Symptom
Possible Cause
Solution
Flipped Image
Reference Length Set Incorrectly
Adjust Reference Length
Scanner(s) Not Scanning (Check
with IR Card in Focus While
Running a Slow 2D Scan)
Scanner Cable Connection Faulty
Switch the System Off and
Reconnect the Electrical
Scanner Cable Between
Base Unit and Scanner
Scanner Calibration Wrong
Please Refer to the
ThorImage
®
OCT Software
Manual for the Scanner
Calibration Procedure
Other Reason
Contact Thorlabs’ OCT
Support
2
Table 4 Troubleshooting