SD-OCT Base Unit
Chapter 4: System Operation
Rev C, June 25, 2021
Page 31
Adjusting the Reference Intensity
For optimum imaging quality it is necessary to ensure that the reference intensity is set within the correct range.
The reference intensity is displayed in ThorImage
®
OCT, as shown below. For optimizing reference intensity,
use the reference intensity adjustment knob (see Figure 24 for the OCTG scanner or the OCTH manual if you
are using a Thorlabs handheld scanner).
Figure 33 Reference Intensity Indicator Bar in ThorImage
®
OCT
Start a B-scan acquisition. If the reference intensity bar is low, turn the knob
clockwise
to optimize the
reference intensity. In case the reference intensity bar is too high, turn the knob
counter-clockwise
. The
reference intensity is at its optimum when the bar is green.
Adjusting the Reference Length
The following steps will guide you to a basic adjustment of the reference length:
-
Start B-scan acquisition. When significantly misadjusted, you will get an incorrect OCT image or no
image of your sample, irrespective of the focus position.
-
Turn the reference length adjustment micrometer screw
counterclockwise
(see Figure 24), until the
force needed for turning significantly increases.
-
Now, slowly turn the reference length adjustment screw
clockwise
until you see the signal from the
sample coming into the displayed B-scan from the bottom.
-
Click the auto-adjust button (see ThorImage
®
OCT Software Manual) to adjust the dynamic range of
your B-scan.
-
When using the IR card for adjustment, your B-scan image should look as shown in Figure 34.
No Reference Intensity
Very Low Reference Intensity
Low Reference Intensity
Optimum Reference Intensity
High Reference Intensity
Very High Reference Intensity