Chapter 7: Troubleshooting
Rev C, June 25, 2021
Page 55
Chapter 7
Troubleshooting
Note:
After switching off the base unit’s power supply with the power switch on the back panel, always wait for
10 seconds before switching it back on again.
2
Please refer to Chapter 13 for Thorlabs OCT Support contact info.
Problem
Possible Cause
Solution
System Does Not Start
No Power Supplied to the Unit
Connect the Power Supply
Power Cord is Broken
Change Power Cord
Other Reason
Contact Thorlabs’ OCT Support
2
Standby Button Does Not
Turn Green when Pushed or
Does Not Light Up at All
Power Switch on Back Panel Set to
Off Position
Switch Power on Using Back
Panel Switch
Interlock Indicator Lights Up
Interlock is Open
Close Interlock Circuit and
Power-Cycle the OCT Base Unit
SLD ON Indicator is Blinking
SLD Power Below Threshold
Contact Thorlabs’ OCT Support
2
Sys OK Indicator is Blinking
Power Supply Defect
Contact Thorlabs’ OCT Support
2
System Does Not Perform
Measurements
PC Crashed
Restart PC
Poor Connection of USB Cable
Check USB Connection
Data Acquisition Cable Not Connected Connect Data Acquisition Cable
Optical Path Length Not Matched
Adjust Optical Path Length
Beam is Blocked
Clean Fiber Tip
Interlock is Open
Close Interlock and Restart
Base Unit
Issue with Camera
Reset Camera in ThorImage
(for CameraLink)
Reset Camera by removing
USB 3.0 Camera cable
(for USB 3.0)
Other Reason
Contact Thorlabs’ OCT Support
2
No Signal in Image
Fiber Not Connected or Not
Connected Correctly
Connect Fiber Patch Cable
Fiber Tip is Dirty
Clean Fiber Tip
Scanner Not Connected
Connect Scanner
Focus is Out of Imaging Area
Adjust Reference Length,
Readjust Focus
Reference Intensity too High or too
Low
Adjust Reference Intensity
Other Reason
Contact Thorlabs OCT Support
2
Bad Image Quality
(See also “No Signal in
Image”)
Under-Sampled, Mirrored Image
Shown
Observe the OCT Image While
Adjusting the Distance of the
Scanner to the Sample
Distance to the Sample is too Short
When Decreasing the Distance,
the Image Needs to Move
Towards the Top of the OCT
Image
Reference Intensity too High or too
Low
Adjust Reference Intensity
Other Reason
Contact Thorlabs’ OCT Support