If the selected Clip Level is too small, this increases the Calculation Area,
captures noise around the beam and this way the measured beam width is
larger. This leads to an increased M².
M² is smaller than 1.0 - How can this be?
M² values < 1.0 are non-physical but may be due to
a too small Calculation Area (Clip Level too high - see
)
the accuracy of the measured result. A error of 5% should be considered.
incorrect wavelength setting. Set the wavelength to the correct value; M²
results will be corected without running a new measurement.
The beam profile looks distorted (particularly, at the end positions of the
stage).
Even if a laser is expected to produce a Gaussian beam with M² = 1.0, the beam still
can be influenced by every optical element between laser and Beam Profiler. For
instance, a focussing lens could be mounted with a tilt or could produce a height
distortion which results in optical aberrations.This reduces the beam quality.
Filters and mirrors may impact on the beam profile as well in the case that surfaces
are contaminated. Clean surfaces according to manufacturer's instructions.
107
Содержание BC106-UV
Страница 1: ...2011 Operation Manual Thorlabs Beam Beam Analyzing Software BC106 VIS BC106 UV ...
Страница 2: ...Version Date 4 0 11 04 2011 2011Thorlabs 2011Thorlabs ...
Страница 76: ... 2011Thorlabs 74 Thorlabs Beam 4 0 Example ...
Страница 157: ... 2011Thorlabs 155 Appendix 7 4 4 Drawings 7 4 4 1 BC106 UV ...
Страница 158: ... 2011Thorlabs 156 Thorlabs Beam 4 0 7 4 4 2 BC106 VIS ...
Страница 159: ... 2011Thorlabs 157 Appendix 7 4 4 3 BC1M2 Mounting Adapter ...
Страница 160: ... 2011Thorlabs 158 Thorlabs Beam 4 0 7 4 4 4 Translation Stage VT 80 ...
Страница 167: ......