Thorlabs Beam 4.0
3.3.2.4.1 Pass/Fail Test
The Calculation Results panel includes a pass/fail test.
For each parameter a minimum and maximum can be set.
Pass/Fail test result will be displayed only for those parameters, where the
appropriate box in Pass/Fail column is checked.
The test result status will be displayed beside the check box verbally and by red/
green color.
Note
Please note the following explanations to the parameter
Ellipse (fitted)
, Diameter (13.5%):
As per definition, the beam ellipse has a minor and a major axis, also known as
minimum and maximum ellipse diameter.
For pass/fail testing, for both diameters can be entered an upper and lower margin.
In the given above example, the "pass" ranges are:
-
Min:
minor axis (
Min
. diameter) must be between 5240 and 5440µm
-
Max:
major axis (
Max
. diameter) must be between 5455 and 5515µm
Only if both conditions are fulfilled, the test has been passed.
Toolbar
Symbol
Associated Action
Save diagram (screenshot) or measurement data (TXT, CSV and
XLS formats available)
Lock or unlock the test parameter
Load the test parameter configuration
Save the test parameter configuration
Содержание BC106-UV
Страница 1: ...2011 Operation Manual Thorlabs Beam Beam Analyzing Software BC106 VIS BC106 UV ...
Страница 2: ...Version Date 4 0 11 04 2011 2011Thorlabs 2011Thorlabs ...
Страница 76: ... 2011Thorlabs 74 Thorlabs Beam 4 0 Example ...
Страница 157: ... 2011Thorlabs 155 Appendix 7 4 4 Drawings 7 4 4 1 BC106 UV ...
Страница 158: ... 2011Thorlabs 156 Thorlabs Beam 4 0 7 4 4 2 BC106 VIS ...
Страница 159: ... 2011Thorlabs 157 Appendix 7 4 4 3 BC1M2 Mounting Adapter ...
Страница 160: ... 2011Thorlabs 158 Thorlabs Beam 4 0 7 4 4 4 Translation Stage VT 80 ...
Страница 167: ......