Page 10 of 18
5.3 Replacing the UVC Lamp
To assure sterilisation of the granulate, replace the UVC lamp after 8000 operating hours. Replacement of the lamp is
shown on the pictures below.
Before replacing the UVC lamp turn the machine off b
y switching the rocker switch to “O” and disconnect the power
plug from the electrical socket.
1. Open the top lid and remove the upper screws
from the front cover using a 4mm allen wrench.
Loosen the bottom 2 screws by turning them counter
clockwise. (1-2 rotations)
2. Open the front cover
3. Slide the lamp carefully away from the socket
4. Remove the lamp
5. Insert the new lamp into the socket by sliding it
frontwards until it stops
6. Close the front cover and top lid. Insert the
screws and fasten them.
UVC-Lamp