8.XXX PARTIAL SHORT-CIRCUIT FACTOR ( A multiplier for the instantaneous standard current is shown on
the display). Press the UP and DOWN keys to change this value.
This function is used to set an instantaneous current threshold, above which a partial short-circuit occurs and the
temperature controller sends fault signal F097.
The standard instantaneous current is input during the calibration procedure with the system hot ( see par. 5.2 )
•
Press the RESET/MODE key.
•
9.XXX FAULT DISPLAY ENABLE (numeric codes appear to enable/disable certain fault messages). Press the UP
and DOWN keys to change this value.
0 = disable all fault messages
255 = enable all fault messages
To enable certain fault messages only, the code is determined as follows:
ENABLE FAULT 82
1 +
ENABLE FAULT 84
2 +
ENABLE FAULT 86
4 +
ENABLE FAULT 87
8 +
ENABLE FAULT 89
16 +
ENABLE FAULT 97
32
CODE DETERMINED = 63
•
A.XXX RATED CURRENT ( the rated current of the temperature controller appears on the display as 030 /
060 ): Press the UP and DOWN keys to change this value.
The rated current depends on the temperature controller hardware.
Note : To increase the level of protection, this parameter is fixed and cannot be changed with the new boards.
•
Press the RESET/MODE key.
•
B.XXX KD GAIN (the value corresponding to the derivative gain appears on the display). Press the UP and DOWN
keys to change this value.
•
Press the RESET/MODE key.
•
C.XXX ENABLE COMPENSATION WITH SYSTEM COLD ( 000/ 001 appears on the display). Press the UP
and DOWN keys to change this value.
This is used to enable the initial pre-heat function , 001=enable; for further information about this function, refer
to paragraph 5.6.
•
D.XXX SERIAL PRINTER ENABLING ( 000/ 001 appears on the display). Press the UP and DOWN keys to
change this value.
This function allows data to be exchanged with the 3E GRAPHIC PRINTER.
•
E.XXX SEALER NO. FOR PRINTER ( the number assigned to the heat-sealer appears on the display).
Press the UP and DOWN keys to change this value.
This function allows the 3E-GRAPHIC PRINTER to recognise the number of the sealer.
•
Press the RESET/MODE key for 3 seconds to exit.
•
F.XXX ENERGY control enabling
•