CONDENSER REQUIREMENTS:
WATER COOLED UNITS
The facility cooling water supply must meet or exceed the re-
quirements shown below for the unit to operate at its full rated
capacity.
As the temperature of the cooling water supply increases, the
required flow and pressure of the cooling water supply in-
creases.
Example: On the HX-150, if the temperature of the cooling
water supply is 65°F (18°C), the flow rate must be at least 1.5
gpm ( 5.7 lpm ), with a pressure differential of at least 3.5 psi
(24 kPa). However, if the temperature of the cooling water
supply rises to 85°F ( 29°C ), the flow rate must be at least 4.0
gpm (15 lpm ), with a pressure differential of at least 10 psi
( 69 kPa).
A water regulating valve, located in TAP WATER line, regulates
the flow rate of the cooling water supply as it enters the unit.
The valve regulates the flow rate based on the heat load. Flow
through the unit stops automatically when the unit is shut off.
HX 100
Required flow - gpm
Required flow - lpm
Required pressure drop - PSI
Required pressure drop - BAR
Required pressure drop - kPa
HX 150
Required flow - gpm
Required flow - lpm
Required pressure drop - PSI
Required pressure drop - BAR
Required pressure drop - kPa
1.0
3.7
2
0.13
13
3.5
13.2
10
0.69
69
2
7.6
5.0
0.34
34
1.5
5.7
3.5
0.24
24
55°F (13°C)
65°F (18°C)
75°F (24°C)
85°F (29°C)
55°F (13°C)
65°F (18°C)
75°F (24°C)
85°F (29°C)
1.0
3.7
2
0.13
13
1.5
5.7
3.5
0.24
24
2.5
9.5
6
0.41
41
4
15.1
10
0.69
69