CONDENSER REQUIREMENTS:
AIR COOLED UNITS
The unit should be located in a laboratory or clean industrial
environment where ambient temperatures are inside the range
of +55°F to +95°F (+13°C to +35°C ). The unit will retain its
full rated capacity in ambient temperatures to approximately
+75°F (+24°C ). Above +75°F, derate the cooling capacity
1% for every 1°F above +75°F, to a maximum ambient
temperature of +95°F. The Celsius deration is 1% for every 0.5
°C above 24 °C, to a maximum ambient temperature of +35°C.
The unit must be positioned so the air intake and discharge are
not impeded. Air is drawn in through the left side of the unit and
discharged through the right and rear. A minimum clearance of
2 feet (0.6 meter) on these three sides is necessary for ad-
equate ventilation. Inadequate ventilation will cause a reduction
in cooling capacity and, in extreme cases, compressor failure.
Excessively dusty areas should be avoided and a periodic
cleaning schedule should be instituted.
Refer to the list below to determine the approximate amount of
air intake required for the unit to retain its full rated capacity.
HX100:
710 cfm (20100 lpm)
HX150:
1050 cfm (29730 lpm)