Leak In Base-Mounted Type
The base mount type may develop a leak across the
diaphragm. Refrigerant will escape into the adjustment
chamber. The refrigerant will be trapped in the adjustment
chamber (if the dust cap is in place). Over time, the adjustment
chamber pressure will rise from ambient to whatever pressure is
present in the refrigeration system. The pressure across the
diaphragm will equalize. The diaphragm in the valve will cease
to operate correctly, as it expects to have ambient air pressure
on one side.
Hot Gas Valve Leak
Leak in Vertical Type
If a leak is suspected, check for leaks carefully where the ad-
justing screw enters the valve body. If an internal leak occurs,
the refrigerant will escape along the threads of the adjusting
screw. This will require replacement of the valve.