C-Link Protocol Commands
Diagnostics
Thermo Electron Corporation
Model 48
i
Instruction Manual
B-15
These commands set the CO concentration alarm minimum
value,
where
value
is set to either floor or ceiling, according to Table B-3. The example
below sets the CO concentration minimum alarm trigger to ceiling.
Send:
set alarm trig conc 1
Receive:
set alarm trig conc 1 ok
Diagnostics
agc int
This command reports the current intensity of the reference channel AGC
circuit. The example below reports that the current AGC intensity is
200,000 Hz.
Send:
agc int
Receive:
agc int 250000 Hz
bias voltage
This command reports the current IR bias supply voltage. The example
below reports that the bias voltage is -102.3 volts
Send:
bias voltage
Receive:
bias voltage -102.3 V
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, positive 3.3, positive 18 IR, positive 18
MOT, and VBIAS. Each voltage value is separated by a space.
Send:
diag volt mib
Receive:
diag volt mib 24.1 14.9 -14.9 4.9 3.2 17.9 17.9
Table B-3.
Alarm Trigger Values
Value
Alarm Trigger
00
Floor
01
Ceiling
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part number 101891 00 8Apr2005 ...
Страница 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Страница 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 236: ...System Description I O Components 8 14 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Страница 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...
Страница 243: ...Thermo Electron Corporation Model 48i Instruction Manual 9 7 Figure 9 5 Retrofit Rack Mounting ...
Страница 244: ...Optional Equipment Mounting Options 9 8 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 292: ...C Link Protocol Commands Record Layout Definition B 46 Model 48i Instruction Manual Thermo Electron Corporation ...