C-Link Protocol Commands
Alarms
B-14
Model 48
i
Instruction Manual
Thermo Electron Corporation
alarm pressure max
These commands report the pressure alarm minimum and maximum value
current settings. The example below reports that the pressure alarm
minimum value is 205 mmHg.
Send:
pressure alarm min
Receive:
pressure alarm min 205 mmHg
set alarm pressure min
value
set alarm pressure max
value
These commands set the pressure alarm minimum and maximum values to
value,
where
value
is a floating-point number representing pressure alarm
limits in millimeters of mercury. The example below sets the pressure alarm
maximum value to 215 mmHg.
Send:
set alarm pressure max 215
Receive:
set alarm pressure max 215 ok
alarm sample flow min
alarm sample flow max
These commands report the sample flow alarm minimum and maximum
value current settings. The example below reports that the sample flow
alarm minimum value is 2 LPM.
Send:
alarm sample flow min
Receive:
alarm sample flow min 2 l/min
set alarm sample flow min
value
set alarm sample flow max
value
These commands set the sample flow alarm minimum and maximum values
to
value,
where
value
is a floating-point number representing sample flow
alarm limits in liters per minute. The example below sets the sample flow
alarm maximum value to 1 LPM.
Send:
set alarm sample flow max 1
Receive:
set alarm sample flow max 1 ok
alarm trig conc
This command reports the CO concentration alarm trigger action for
minimum alarm, current setting, to either floor or ceiling. The example
below shows the CO concentration minimum alarm trigger to ceiling,
according to Table B-3.
Send:
alarm trig conc
Receive:
alarm trig conc 1
set alarm trig conc
value
Содержание 48i
Страница 1: ...Model 48i Instruction Manual Gas Filter Correlation CO Analyzer Part number 101891 00 8Apr2005 ...
Страница 12: ...Preface Thermo Electron Corporation Model 48i Instruction Manual xii ...
Страница 22: ...Installation Startup 2 6 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 222: ...Servicing Service Locations 7 40 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 236: ...System Description I O Components 8 14 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 241: ...Thermo Electron Corporation Model 48i Instruction Manual 9 5 Figure 9 3 Bench Mounting ...
Страница 242: ...9 6 Model 48i Instruction Manual Thermo Electron Corporation Figure 9 4 EIA Rack Mounting ...
Страница 243: ...Thermo Electron Corporation Model 48i Instruction Manual 9 7 Figure 9 5 Retrofit Rack Mounting ...
Страница 244: ...Optional Equipment Mounting Options 9 8 Model 48i Instruction Manual Thermo Electron Corporation ...
Страница 292: ...C Link Protocol Commands Record Layout Definition B 46 Model 48i Instruction Manual Thermo Electron Corporation ...